ApplicationWheel Tracking Apparatus is suitable for the research, production and quality inspection
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
Comparative tracking index tester is suitable for the research, production and quality inspection departments in the industry of lighting equipment, low-voltage electrical appliances, household appliances, machine tools, electrical machine, electric tools, electronic instruments, electro technical instruments, information technology equipment, as well as insulating materials , engineering plastics, electrical connectors, accessories industry.
GBT 4207.IEC 60112 (2003),IEC 60745-1.IEC 60335-1.IEC 60598-1.IEC 60884-1EC60112
1. The chamber is paint spray procced, elegant and anti-corrosive, size:1200mm X 600mmx 1200mm.
2. All specimen holders are made by stainless steel material.
3. Equipment consists of the voltmeter, ammeter, delay overcurrent relay, power supply, circuit control system, button switch, signal lights and other components.
4. Automatic drip device contains drip tube, drip nozzle, counting device, drip counting device, drip control device, droplet volume measuring tool, etc.
5. Drip nozzle device can adjust; Adjust height 30 to 40 mm, ensure the drip position between the two electrodes .
6. Electrode device contains platinum electrode, electrode fixing device, 4 mm gauge, etc. The electrode is adjustable and detachable.
7. Electrode is 5 mm wide, 2 mm thick, 99.9% purity of platinum, 30 ~ 60 degrees sharp tip.
8. Electrode pressure to the sample can be adjusted. The adjusting range is 1.00 N (+ / - 0.001 N), electrode distance 4.0 mm + / - 0.01 mm, intersection Angle 60 ° + / - 5 ° .
9. Equipped with sample stage
10. Equipped with glass plate: thickness≥4mm
11. Equipped with electrolyte:1 bottle of ammonium chloride and 1 bottle of alkyl naphthalene sulfonic acid sodium reagent.
12. Test voltage: 0 ~ 1000 v adjustable and digital display. Maximum error of voltage meter display value is 1%.
13. Short circuit current: (1 ±0. 1), A voltage drop value less than 10% .
14. Test current over 0.5 A lasting more than 2 seconds, the system automatically shut down the equipment.
15. Safety door control system: when the door is opened, the test system will not work.
16. Electrode: 5 mm x 2 mm rectangle platinum electrode and brass electrode each one pair, in accordance with the size requirements of IEC60112 figure B. 1.
17. The intersection angle between electrodes is 60 °±5 °, spacing of 4±0.1 mm.
18. Electrode pressure to sample is 1. 00N±0.05N.
Controller | Button control / touch screen control |
Voltage Display | Digital display |
Current Display | Digital display |
Dimensions | 1120 x 600 x 1020(L x W x H) |
Inner Capacity | ≥0.5cube |
Electrode Material | Pt |
Electrode Size | 2mm x 5mm x 35 mm |
Electrode Pressure | 1.00N |
Electrode Distance | 4.0mm |
Droplet Volume | 50 drop/1cm3 |
Droplet Height | 35mm(can be set) |
Droplet Time | 30s±0.1(can be set) |
Disclaimer: Le contenu ci-dessus est destiné à des fins de référence et de communication uniquement entre les initiés de l'industrie, et ne garantit pas son exactitude ou son exhaustivité. Conformément aux lois et règlements pertinents et aux règlements de ce site Web, les unités ou personnes qui achètent des articles connexes doivent obtenir des qualifications et des conditions de qualification valides.