Catheter Residual Vacuum Tester - Qinsun
Fabricant professionnel d'instruments de détection
Ligne directrice nationale de consultation :+86 021-64208466

Catheter Residual Vacuum Tester

Main Uses of the Catheter Residual Vacuum TesterThe catheter residual vacuum tester is mainly used t

Courriel : info@standard-groups.com

Contenu

Main Uses of the Catheter Residual Vacuum Tester

The catheter residual vacuum tester is mainly used to determine the physical properties of residual vacuum and leakage in suction catheters for the respiratory tract and other similar medical devices. It is one of the important means to evaluate the performance of medical devices such as suction catheters for the respiratory tract, ensuring the safety and effectiveness of these devices during use. By testing the residual vacuum of catheters, the effectiveness of the vacuum control device at the machine end in releasing vacuum can be verified, thus ensuring that respiratory catheters can meet specific performance requirements during use. In addition, this tester can also be used for detecting vacuum leakage in catheter systems of medical devices such as hemodialysis machines, intravenous infusion devices, and ventilators, to promptly identify and resolve leakage issues and avoid medical accidents caused by leakage.


Compliance Standards

YY/T 0339 - 2019 *Suction Catheters for the Respiratory Tract*, etc.


Paramètres techniques

- **Flowmeter**:

   - **Measurement Range**: Usually, it can measure 30L/min or more.    - **Accuracy**: ±5% or higher.    - **Airflow Resistance at a Specific Flow Rate (e.g., 30L/min)**: It should be less than 0.1kPa. - **Manometer**:    - **Measurement Range**: Usually covers ±1kPa or a larger range.    - **Accuracy**: Usually within ±0.01kPa (0.1cmH2O). - **Vacuum Pump**: It is used to create a vacuum environment inside the catheter, and its performance should be stable and reliable. - **Control Device**: It is used to control the operation of the tester and the setting of test parameters, and should have an intuitive and easy - to - use operation interface.


企业微信截图_1741844310880.png

Disclaimer: Le contenu ci-dessus est destiné à des fins de référence et de communication uniquement entre les initiés de l'industrie, et ne garantit pas son exactitude ou son exhaustivité. Conformément aux lois et règlements pertinents et aux règlements de ce site Web, les unités ou personnes qui achètent des articles connexes doivent obtenir des qualifications et des conditions de qualification valides.


Message en ligne

Détails de contact

Téléphone de l'entreprise

+86-21-6420 0566

Heures de travail

Du lundi au vendredi

Téléphone portable:

13816217984

Courriel :

info@qinsun-lab.com

Code QR