- Evaluating the wear - resistance of hip joint prostheses: By simulating human movement states such
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
Standard
Paramètres | Spécification |
---|---|
Angular Motion Range | Flexion/Extension: ±100° Abduction/Adduction: ±30° Rotation: ±40° |
Max. Axial Force | ±4,500 N (1,000 N resolution) |
Load Control Accuracy | ±0.5% of reading (0–4,500 N) |
Frequency Range | 0.1–5 Hz (programmable waveforms) |
Contrôle de la température | 37±2°C (PT1000 sensors) |
Lubrication System | 15L capacity with protein skimmer and bubble trap |
Data Acquisition | 100 kHz sampling rate (16-bit resolution) |
Power Requirements | 220V/110V ±10%, 50/60Hz, 2,500W |
Dimensions (W×D×H) | 1,800 × 1,200 × 2,000 mm |
Weight | 950 kg (including bath and control cabinet) |
Noise Level | <65 dB (A-weighted) |
Disclaimer: Le contenu ci-dessus est destiné à des fins de référence et de communication uniquement entre les initiés de l'industrie, et ne garantit pas son exactitude ou son exhaustivité. Conformément aux lois et règlements pertinents et aux règlements de ce site Web, les unités ou personnes qui achètent des articles connexes doivent obtenir des qualifications et des conditions de qualification valides.
Téléphone de l'entreprise
+86-21-6420 0566
Heures de travail
Du lundi au vendredi
Téléphone portable:
13816217984