Medical mask bacterial filtration efficiency (BFE) tester

It is composed of aerosol generation system, aerosol chamber and aerosol transmission device, negative pressure cabinet and 28.3l /min sampler.The whole tester is controlled by the console.The control console adopts microcomputer control system to coordinate and control the work of aerosol generation system, transmission system, negative pressure cabinet and sampling system, and display the working status in real time. The whole measurement work is completed automatically.

Mask Bacterial Filtration Efficiency Tester Features
 

Double six-stage Anderson microbial sampler test systems with contrast sampling tests provide high accuracy.
Internal aerosol generator provides stable and adjustable microbial aerosol flow. Internal pump can adjust the flow rate.
Equipped negative pressure test chamber to ensure the safety of operators. Equipped light system to make the operations more clear and easy to do.
10.4 Inch touch screen control panel is easy to operate and supports USB data transmission.
Equipped leakage protection and operating window provide better operating view and safety operating conditions.
Stainless steel chamber with insulation and fire retardant lining is stable and reliable.


Testeur d'efficacité de filtration bactérienne BFE Application:
 
Mask Bacterial Filtration Efficiency Tester is applied to test the filtration efficiency of mask, can be used in metering calibration department, scientific research institution, mask manufacturer and other related departments.

Specification:


Key SpecificationSpecification RangeResolutionMax. permissible errors
A Route Sampling Flow28.3L/min0.1L/minWithin ± 2.0%
B Route Sampling Flow28.3L/min0.1L/minWithin ±2.0%
Spray Flow(8~10)L/min0.1L/minWithin ±5.0%
Peristaltic Pump Flow(0.006~3.0)mL/min0.001ml/minWithin ±2.0%
Front Pressure of A Route Sampling Flowmeter(-20~0)kPa0.01kPaWithin ±2.0%
Front Pressure of B Route Sampling Flowmeter(-20~0)kPa0.01kPaWithin ±2.0%
Front Pressure of Spray  Flowmeter(0~300)kPa0.1kPaWithin ±2.0%
Negative Pressure of the Chamber(-90~-120)Pa0.1PaWithin ±1.0%
Working Temperature0-50℃
Negative Pressure of the Chamber>120Pa
Data Save Ability>500000 sets(Scalable capacity)

Medical mask bacterial filtration efficiency (BFE) tester

Disclaimer: Le contenu ci-dessus est destiné à des fins de référence et de communication uniquement entre les initiés de l'industrie, et ne garantit pas son exactitude ou son exhaustivité. Conformément aux lois et règlements pertinents et aux règlements de ce site Web, les unités ou personnes qui achètent des articles connexes doivent obtenir des qualifications et des conditions de qualification valides.


Message en ligne

Détails de contact

Téléphone de l'entreprise

+86-21-6420 0566

Heures de travail

Du lundi au vendredi

Téléphone portable:

13816217984

Courriel :

info@qinsun-lab.com

Code QR