The filter measuring device consists of 2 components: Filter measuring device FMP 03 and universal f
Courriel : info@standard-groups.com
The filter measuring device consists of 2 components: Filter measuring device FMP 03 and universal filter holder FH-143/149 automatic. The automatic measuring set FMP03 + FH is suitable for quickly testing particle filters and filter media.
Application
For production controls and type testing, e.g. testing filter respirators and filtering half-mask face pieces with liquid paraffin aerosol according to standards EN 143 and EN 149.
Leak testing and filter resistance tests on particle filters to a maximum diameter of 200 mm. Very good correlation to corresponding measurements according to EN 1822.
Features
· Short testing times (approx. 10 seconds per filter) and extremely simple operation
· The separation efficiency and filter resistance of the sample are determined in a sample clamp.
· The plant process control computer carries out the test on its own according to the user's specifications.
· The test is carried out automatically according to the specified test standards, either with a fixed or different volumetric flows, with uniform or different aerosol concentrations or with uniform or different flow directions.
· The process control computer assesses the sample.
· The only thing the operator has to do is insert and remove the sample.
Disclaimer: Le contenu ci-dessus est destiné à des fins de référence et de communication uniquement entre les initiés de l'industrie, et ne garantit pas son exactitude ou son exhaustivité. Conformément aux lois et règlements pertinents et aux règlements de ce site Web, les unités ou personnes qui achètent des articles connexes doivent obtenir des qualifications et des conditions de qualification valides.
Téléphone de l'entreprise
+86-21-6420 0566
Heures de travail
Du lundi au vendredi
Téléphone portable:
13816217984
Courriel :
info@qinsun-lab.com