StandardASTM F2003 (Standard Test Method for Fatigue Testing of Orthopedic Implants)ISO 14879-1 (Tot
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
YY/T 0809.4 (China): Specifies fatigue testing protocols for femoral stems in partial/total hip arthroplasty.
ISO 7206-4 (International): Defines methodologies for evaluating the fatigue strength of femoral stem-bone cement interfaces.
ASTM F2580: Additional compliance for impact resistance testing of modular femoral necks.
Adaptive Protocols: Customizable for emerging standards in additive-manufactured implant evaluation.
Electromagnetic Precision: M-Series actuator delivers ±0.05% force accuracy with 0.001N resolution.
5 Million Cycle Endurance: Proven capability to complete ISO 7206-4’s 5×10⁶-cycle fatigue test without drift.
R=0.1 Load Control: Advanced closed-loop system maintains stress ratios during dynamic loading.
Real-Time Fracture Detection: AI-powered acoustic emission monitoring identifies micro-cracks at sub-millisecond intervals.
Intuitive Workflow: 21.5” capacitive touchscreen with gesture control and pre-loaded SOPs for 15+ implant geometries.
Modular Environmental Chambers: Quick-swap pods for simulated physiological (-20°C to +50°C) and corrosive environment testing.
Electromagnetic Actuator: 2.300 N peak force capacity with programmable load ratios (R=0.1–10).
Triaxial Loading System: Simultaneous axial/torsional/shear loading (50 Nm max torque).
Dynamic Load Control: 1–30 Hz frequency range with programmable waveforms (sine, square, random, custom).
Thermal-Mechanical Chamber: ±0.3°C temperature stability with integrated humidity control (10–95% RH).
High-Speed Imaging Ready: Syncs with external cameras for full-field strain mapping (DIC optional upgrade).
Safety First: Triple-redundant emergency stop, overload clutches, and laser-based specimen breakage detection.
Paramètres | Spécification |
---|---|
Max. Dynamic Force | ±2.300 N (5.000 N peak optional) |
Frequency Range | 1–30 Hz (0.1 Hz increments) |
Load Ratio (R) | 0.1–10 (adjustable) |
Axial Displacement | ±15 mm (0.1 μm resolution) |
Torque Capacity | ±50 Nm (0.01 Nm resolution) |
Temperature Range | -20°C to +50°C (environmental chamber) |
Humidity Control | 10–95% RH (non-condensing) |
Taux d'échantillonnage des données | 100 kHz (synchronized across 16 channels) |
Power Requirements | 220V/110V ±10%, 50/60Hz, 2.000W |
Dimensions (W×D×H) | 1.500 × 1.000 × 2.200 mm |
Weight | 750 kg (including environmental chamber) |
Noise Level | <68 dB (A-weighted) |
Femoral Stem Test Fixture Kit:
Self-centering collet system (12/14 mm taper compatibility)
Laser-aligned loading anvils with diamond-like carbon (DLC) coating
Thermal-Mechanical Chamber:
25L capacity with rapid cooling/heating (15°C/min)
316L stainless steel interior with HEPA filtration
TestMaster Ultimate Software:
Automated fatigue life prediction (S-N/ε-N curve generation)
21 CFR Part 11-compliant audit trail with blockchain timestamping
Remote monitoring via dedicated iOS/Android app
Operator Kit:
Calibrated torque wrench set (0.5–50 Nm)
Specimen alignment gauges (ISO 7206-4 compliant)
Calibration Bundle:
NIST-traceable force calibration weights
Temperature/humidity verification probe
Safety Kit:
Laser safety curtains
Acoustic enclosure with 30 dB noise reduction
Disclaimer: Le contenu ci-dessus est destiné à des fins de référence et de communication uniquement entre les initiés de l'industrie, et ne garantit pas son exactitude ou son exhaustivité. Conformément aux lois et règlements pertinents et aux règlements de ce site Web, les unités ou personnes qui achètent des articles connexes doivent obtenir des qualifications et des conditions de qualification valides.