The next generation impactor TM, or NGI TM, is a high performance cascade impactor for classifying a
Courriel : info@standard-groups.com
The next generation impactor TM, or NGI TM, is a high performance cascade impactor for classifying aerosol particle into size fractions for testing meter-dose and dry-power inhaler and other inhaled drug delivery devices such as nebulizers and nasal sprays. MSP developed the NGI in conjunction with a consortium of 15 pharmaceutical companies, resulting in superior aerodynamic performance, low inter-stage losses, and user-friendly features, The NGI is recognized internationally both in the USP and the EP. It features removable collection cups, O-ring-free connections, and a micro-orifice collector that can be used in place of a final filter for most formulation. These features make the NGI an easy-to-use, efficient workstation, enabling 10 to 20 inhaler determinations per day with one impactor, resulting in less capital equipment and a cleaner, more efficient laboratory.
Designed specifically for the pharmaceutical industry for testing metered-dose inhalers (MDI), dry-power inhalers (DPI), nebulizer and nasal sprays.
Provides high-accuracy data and unappalled productivity.
Features
1. Seven impaction stages plus a final micro-orifice collector that can be used in place of a final filter for most formulations.
2. Known stage cut size from 15L/min to 100L/min with no change of parts needed for entre range.
3. Also calibrated at 15L/min specifically for nebulizers.
4. Seven stages with at least five with D50 values between 0.54 and 6.12 μm at all design flow rates.
5. Removable set of collection cups, hinged removable lid, and O-ring free inlet connections.
6. Electrically conductive and corrosion resistant.
7. 316 stainless steel USP induction port.
8. One pre-separator for entire range of 30L/min to 100L/min W/O changing parts.
Technical Parameters
Projets | Paramètres |
Flow rate | 15-100L/min |
Stage cut point(1-7 stages) | Look up the USP 601 Section, Table 6 for more information |
D50 @30L/min | 11.4, 6.4, 4.0, 2.3, 1.4, 0.80, 0.54 µm |
D50 @ 60 L/min | 8.1, 4.4, 2.8, 1.7, 0.94, 0.55, 0.34 µm |
D50 @100 L/min | 6.1, 3.4, 2.2, 1.3, 0.72, 0.40, 0.24 µm |
Weight | 9kg |
Dimension | 600 x 330 x 130 mm (L x W x H) |
Disclaimer: Le contenu ci-dessus est destiné à des fins de référence et de communication uniquement entre les initiés de l'industrie, et ne garantit pas son exactitude ou son exhaustivité. Conformément aux lois et règlements pertinents et aux règlements de ce site Web, les unités ou personnes qui achètent des articles connexes doivent obtenir des qualifications et des conditions de qualification valides.