Simple Porometer - Qinsun
Fabricant professionnel d'instruments de détection
Ligne directrice nationale de consultation :+86 021-64208466

Simple Porometer

Principle A wetting liquid is allowed to spontaneously fill the pores in the sample and a nonreactin

Courriel : info@standard-groups.com

Contenu

Principle
 A wetting liquid is allowed to spontaneously fill the pores in the sample and a nonreacting gas is allowed to displace liquid from the pores. The gas pressure and flow rates through wet and dry samples are accurately measured. The gas pressure required to remove liquid from the pores and cause gas to flow is given by
D = 4 γ cos θ / p
where D is the pore diameter, γ is the surface tension of liquid, θ is the contact angle of liquid, and p is the differential gas pressure. From measured gas pressure and flow rates, the pore throat diameters, pore size distribution, and gas permeability are calculated.
 
Application
 The unique features of the instrument, especially its ability to generate highly reproducible data quickly, make it highly suitable for quality control and process control operations. The instrument is used in these applications in a wide variety of industries.

Capabilities
 ● Diameter of the most constricted part of a through pore (pore throat)
● Mean flow pore diameter (50% of flow is through pores smaller than the mean flow pore)
● Bubble Point (the largest through pore throat diameter)
● Pore diameter range
● Pore distribution:
· f = -d[(fw/fd) x 100] / dD
· fw = flow rate through wet sample
· fd = flow rate through dry sample
● Gas permeability in many desired units including Frazier, Gurley, Rayl & Darcy
 
Features
 ◆ Fully Automated and User-Friendly
◆ Only a few minutes per test
◆ Preset test parameters
◆ Very little operator time and involvement 
◆ Ideal for quick generation of highly reproducible data
◆ Variety of sample shapes and sizes
◆ Very little maintenance
◆ Real-time graphical display
◆ Windows-based software for all control, measurement, data collection, data reduction, and report preparation
 
Paramètres techniques

ProjetsParamètres
Pressure Accuracy0.15% of reading
Test Pressure100 psi instrument version (700kPa instrument versions)
Pressure & Flow Resolution1/60,000 of full scale (1 part in 60,000)
Maximum Pore Size Detectable80 µm
Minimum Pore Size DetectableVaries with intrusion liquid
Flow RatesUp to 200 SLPM (liters per minute)
Sample Sizes:Standard: 0.1 inch diameter (up to 1.5" thick)
Sample GeometrySheets, Rods, Tubes, Hollow Fibers, Cartridges, Powders
Pore Size Range80 µm to 0.06 µm

 image.png

Disclaimer: Le contenu ci-dessus est destiné à des fins de référence et de communication uniquement entre les initiés de l'industrie, et ne garantit pas son exactitude ou son exhaustivité. Conformément aux lois et règlements pertinents et aux règlements de ce site Web, les unités ou personnes qui achètent des articles connexes doivent obtenir des qualifications et des conditions de qualification valides.